Table 1. HPLC analytical conditions for maysin isolated from Eremochloa ophiuroides

Parameter Condition
Instrument Shimadzu LC-20AD
Detector Shimadzu SPD-20A Prominence UV/VIS Detector
Wavelength 360 nm
Column YMC-Pack C18 Column (4.6 mm I.D. × 150 mm; particle size 5 μm)
Oven temperature 40°C
Flow rate 1.0 mL/min
Injection volume 5 μL
Mobile phase Time A: 0.1% formic acid in H2O B: acetonitrile
5 85 15
10 70 30
15 50 50
20 30 70
25 0 100
30 0 100
31 95 5
40 95 5