Table 2. Validation of analytical method for pantothenic acid
Vitamin | Sensitivity | Precision | Recovery and accuracy |
LOD1) | LOQ2) | Repeatability3) | Reproducibility4) | Reference value6) | Analysis value7) | Recovery (%) | RSD (%) |
(μg/mL) | RSD (%)5) | RSD (%) |
Pantothenic acid | 0.001 | 0.003 | 4.831±0.0028 | 4.832±0.096 | 150±12 | 150.75±0.95 | 100.5 | 0.63 |
LOD, limit of detection; LOD=3×signal/noise.
LOQ, limit of quantification; LOQ=10×signal/noise.
Repeatability refers to the results of 5 independent determinations carried out for the same sample on the same day.
Reproducibility refers to the results of 5 independent determinations carried out on a sample by analyzing 3 replicates of the sample at each day for 5 days.
Relative standard deviation (%).
Reference value is the true value provided by NIST.
Analysis value is the experimental value obtained by HPLC assay for pantothenic acid.