Instrument | Agilent Technologies 1260 infinity | |
---|---|---|
Column | shodex HILICpak VN-50 4D | |
Mobile phase | Gradient conditon | |
Time table (min) | Acetonitrile (%) | |
0 | 89 | |
25 | 89 | |
30 | 81 | |
40 | 80 | |
75 | 75 | |
80 | 73 | |
80.1 | 89 | |
85 | 89 | |
Flow rate | 1.0 mL/min | |
Temperature | 65℃ | |
Injection volume | 10 μL | |
Detector | ELSD | |
Temperature | 90℃ | |
Gain | 10 |