Table 1
HPLC and HPLC/MS condition

HPLC Condition
Instrument Dionex system
Column Inertsil ODS-2(C18) Column (4.6 mm×250 mm; GL Science, Tokyo, Japan)
Solvent A: Deionised water
B: 20% acetonitriles
0-8 min increased 1-99% B
18-29 min kept 99% B
29-32 min decreased 99-1% B
32-40 min kept 1% B
Column temperature 35℃
Flow rate 1 mL/min
Infection volume 20 μL
Mass system
Polarity ESI+
Capillary temperature 275
Voltage Pray: 5 kV and capillary: 15 V
Mass scan range (m/z) 100-600