Table 2. HPLC conditions for analysis of DNJ in mulberry leaf and tea

Item Operation condition
HPLC system Waters e2690/5 HPLC system equipped with 2998 photodiode array detector and autosampler
Column YMC-Pack Pro C18(5 μm, 4.6×250 mm, YMC Inc., Allentown, PA, USA)
Flow rate 1.0 mL/min
Detector UV254 nm
Gradient elution
  Time (min) H2O 100% MeOH  
0 45 55
16 45 55
19 10 90
22 10 90
25 45 55
30 45 55